Saving Changed Settings

When you are satisfied with the settings you have input, select Store to Flash to save your settings. The yellow warning banner should disappear.

If you do not want to save settings, the meter will use the settings shown under Current Ram Settings and the yellow warning Setup Not Stored will remain visible.

Example Settings
Set Avg/Data Rate = 55
Set Number of Samples = 60
Set Number of Cycles
= 24
Set Time Interval
= 010000
The instrument will collect 60 samples, approximately one second apart, once every hour for 24 hours. (See more examples)

 




retrieving and